Best Practices
for the use of
Bed Occupancy Sensors (BOS) and
Chair Occupancy Sensors (COS)
To clean the sensors wipe with a clean, dry cloth. Avoid holding any sensor by its
connection wire while wiping.
Test for correct activation of the BOS or COS using a resistance meter like a FLUKE
Model 175. When not activated, the resistance should be greater than 10 Mohms, when
activated, the resistance should be less than 10 ohms.
In the pictured test the resistance meter is reading beyond its capability to record a
resistance reading. Interpret this as much greater than 10 Mohms.
Seitenansicht 0
1 2 3 4

Inhaltsverzeichnis

Seite 1 - Best Practices

Best Practices for the use of Bed Occupancy Sensors (BOS) and Chair Occupancy Sensors (COS) To clean the sensors wipe with a clean, dry cloth. Avoid

Seite 2

When the sensor is activated the reading should be less than 10 ohms. In this picture the resistance indicated as the sensor is activated is 0.5 ohm

Seite 3

The correct activation force may be tested by placing a five (5) pound weight on to the center of the BOS. The weight should have a pressure foot of

Seite 4

When placing the sensor be sure to have the wire exit the bed opposite the patient bed exit. Shown in this photo is the WRONG orientation for the bed